Turbocharge Your Sequencing
DNBSEQ-T7 Genetic Sequencer




DNBSEQ-T7 can generate 1-7Tb of high-quality data per day for a wide range of applications, including whole genome, deep exome, epigenome, transcriptome, and targeted panel sequencing projects.
Powered by DNBSEQTM Technology, DNBSEQ-T7 makes sequencing more efficient and more productive with advances in biochemical, fluidics, and optical systems.
Low Cost
- <$2.00 per Gb and <$200 per genome
High Accuracy
- Reduced index hopping, no clonal errors
High Speed
- 24-30 hours for PE150
High Flexibility
- Run 4 independent flow cells at any time
Ultra-high Throughput
- 20,000 of 30x WGS in one year
Performance Parameters
# of flow cell | Reads | Read Length | Data output | Run Time | Q30 |
---|---|---|---|---|---|
1-4 | 5800M* | SE50 |
290~1160 Gb |
|
>85% |
PE100 |
1 ~ 5Tb |
|
>85% | ||
PE150 |
1 ~ 7 Tb |
|
>80% |
* The maximum number of effective reads is based on the sequencing of an internal standard library. Actual output may vary with sample types and library preparation methods.
** Run time includes DNB loading, sequencing, and base calling. (need to check if it includes washing time)
The percentage of base above Q30 is the average of an internal standard library over the entire run. The actual performance is affected by factors such as sample type, library quality, and insert fragment length.
Instrument Specifications
Catalog # | 900-000128-00 | |
---|---|---|
Dimensions / Net Weight | 65 in x 35.6 in x 71.5 in | 1687lbs (765 Kg) |
Power | Type | 200~240 V, 50/60 Hz, 30 A |
Rated Power | 3000 VA | |
Operating Environment Requirements | Temperature | 19~25 °C, <2 °C change per hour |
Relative Humidity | 30%RH ~ 80%RH, non-condensing | |
Atmospheric Pressure | 80 kPa~106 kPa | |
Waterproof Rating | IPX0 | |
Altitude | <2000 meters | |
Floor Bearing Capacity | ≥650 Kg/m2 | |
Control Computer Configurations | CPU | Intel CORE I7-7700 4Core x2 3.6GHz |
Internal Storage | 16 Gb RAM | |
HDD | 1 Tb | |
SSD | 128 Gb | |
Operating System | Windows 10 | |
*Support the computer configurations and system updates | ||
Bandwidth for Network Connection | 300 MB/s | For local storage network uploads |
1000 MB/s | For FASTQ computing uploads |
* This model is for Research Use Only. Please refer to the site prep guide for more details.
Your Data Security is Our Top Priority.
All our sequencing platforms can work as standalone systems from sample to result without any network connection. As a result, there is no risk of a data breach and you can be assured your data is secure.
Disclaimer:
Unless otherwise informed, StandardMPS and CoolMPS sequencing reagents, and sequencers for use with such reagents are not available in Germany, Spain, UK, Hong Kong, Sweden, Belgium, Italy, Finland, Czech Republic, Switzerland, Portugal, Austria and Romania. No purchase orders for StandardMPS products will be accepted in the USA until after January 1, 2023.

2904 Orchard Parkway, San Jose, California 95134
*All products shown are for research use only. Not for use in diagnostic procedures.
Copyright © 2013 - 2022 Complete Genomics Incorporated. All rights reserved. Use of this website signifies your agreement to the Terms of Use and Online Privacy Policy.